MA5690-2 Data Acquisition System

9-99 Measuring Inputs

DESCRIPTION

Model MA5690-2 data acquisition system with graphics display accepts 9 to 99 measuring inputs depending on the configuration ordered. Operation is easy by means of 4 soft keys and cursor block, comprehensive menu system with wizards and help windows.

Technical Features

- Ideal display with large, brightly illuminated 
     graphics
- Easy and convenient operation by means of 
     4 soft-keys and cursor block,
     comprehensive menu system with wizards and help windows
- Choice of languages : German, English, French (others also available)
- 9 measuring menus (3 can be freely configured by user from a range 
     of 50 functions)
- Measured values can be displayed graphically in line chart or bar chart form
     and from 1 to 20 measured values can be displayed numerically in various
     sizes.
- 8 programming menus for easy-to-understand parametrization of cycles, 
     times, memory, device locking, output modules, and power supply, Sensor
     programming with range, units, comments, scaling, error correction, etc.
- 9 LEDs for indicating various operating states
- Data logger with multimedia card, provided as standard, recording in standard
     FAT16 file format, transmission to PC via USB card reader
- Option EEPROM with capacity for 100,000 measures values, internally
     configurable as linear or ring memory
- Sleep mode for long-term recording
- Measuring functions Measured value, zero-setting, setpoint adjustment, 
     maximum and minimum values stored with date and time-of-day, smoothing,
     average values over time or over measuring points, limit value monitoring, 
     cold junction compensation, temperature compensation, and atmospheric
     pressure compensation
- Sensor programming on the ALMEMO® connector itself
- 9 ALMEMO® input sockets, electrically isolated, for 9 ALMEMO®
     sensors, 31additional channels (maximum), (of which 4 internal 
     function channels)
- Extendible up to 99 inputs with various selector switch boards and maximum 
     100 meas. channels
- High-speed, high-resolution A/D converter, 24-bit, 2.5 to 100 measuring
     operations per second
- Electr. isol. between measuring inputs and supply
- 2 ALMEMO® output sockets for digital interfaces, analog outputs, trigger input, 
     and alarm contacts
- 5 LEDs for indicating various operating states in the measuring circuit
- Cold junction measuring with 2 NTCs / insert and interpolation Ranges Pt100,
     0.000 to 65.000 °C, timer 6500.0 seconds

SPECIFICATIONS

Standard equipment

Display: Graphics 128 x 128 dots, 16 rows, Lighting: 5 LEDs, 3 brightness levels
Operation: 9 keys (4 soft-keys and cursor block), 9 status LEDs on the front panel
Memory: Multimedia card and USB card reader
Memory (option S): Internal 512-KB EEPROM (100,000 measured values)
Date and time-of-day: Real-time clock, buffered with lithium battery

Measuring inputs

Master meas. board: 9 ALMEMO®input sockets, electr. isol., with 
   semiconductor relays
Channels: 9 primary channels, 27 additional channels for double sensors 
   and function channels, 4 internal function channels
A/D converter: delta-sigma, 24-bit, 2.5 / 10 / 50 / 100 measuring operations per
   second, electrically isolated
Sensor power supply: AC adapter, 12 V, maximum 400 mA, Rechargeable battery,
   9 to 11 V, maximum 200 mA

Selector switch boards

ES 5690-UA10: 10 electrically isolated inputs for ALMEMO® flat connectors, 10 to 
   40 channels, with sensor supply, 2 plug-in slots
ES 5690-UMU: 10 electrically isolated inputs, sensor connection with 10x MU 
   connector, 10 to 40 channels, without sensor supply, ranges only, for all 
   thermocouples, Pt100, Ni100, NTC, ohms, 2.6 V, 260 mV, 55 mV, 26 mV, 
   1 plug-in slot
ES 5690-UTH: 10 electr. isol. inputs for all thermocouples with miniature thermal 
   connector, 10 to 40 channels,1+1 plug-in slots (dummy panel needed)
ES 5690-UKS: 10 electr. isol. inputs, sensor connection with clamp connector, 
   10 to 40 channels, without sensor supply, ranges only, Pt100, Ni100, NTC, 
   ohms, 2.6 V, 260 mV, 55 mV, 26 mV, 1 plug-in slot
OA 5690-UKSU: Option for ES5690-UKS, inputs with 100:1 divider for 10 V
OA 5690-UKSI: Option for ES5690-UKS, inputs with shunt for 20 mA ranges mA 
   and (4-20 mA)
Dimensions: 
   TG 32 DU 179 x 158 x 232 mm (WxHxD) 6 slots 
   TG 84 DU 444 x 158 x 232 mm (WxHxD) 19 slots 
   BT 84 DU 483 x 132 x 273 mm (WxHxD) 19 slots 
Outputs: 2 ALMEMO®-sockets for all output modules (analog, data, trigger, relais cable etc.) 
   internal alarm transmitter
Power supply: AC adapter: ZB1212NA3/USA 90..250V, 12V DC, 2A; DC adapter 
   cable, electr. isolated ZB3090-UK2, 10..30V,12V DC ,1.25A; Rechargeable 
   battery pack 8 NiMH-Akkus 9..11V, 1600mAh
MA5690-2 Data Acquisition System
(1) LCD
Status bar 
     Cont. Measuring point scan
     Start / stop measuring
     REC Record to memory
     COM Measured value output
     l>, >l Program start / end of measuring
     R01 Status of alarm relays
     * , * Lighting on, pause
     Battery operation / charge status
     13 lines for functions
     Function of keys F1, F2, F3, F4
(2) Check lamps
     START Measuring operation started
     REC Measuring with results saved
     COM Measuring with output
     AVG Averaging
     ALARM Limit value exceeded
           Sensor breakage, LoBat
     LOCKED Keys locked
     SLEEP Flashes in sleep mode.
     CHARGE Battery is being charged.
     Goes out as soon as fully charged.
(3) Keypad
     F1 to F4 Function keys (soft keys)
            Cursor block
     ON PROG Switch on Program, Switch OFF (press and hold down)
     Arrow Keys: Function selection, input
(4) Slot, multimedia card: MMC slot for multimedia memory card
MA5690-2 Data Acquisition System
(5) Module AP rechargeable battery (option)
(a) Connection socket DC-A 12V
      AC adapter (ZB 1212-NA3/USA, 12V, 2A)
(b) Check lamps
      DC-A Mains supply present
      CHARGE Battery is being charged. 
      Goes out as soon as fully charged.
(6) Module MM-A9 Measuring circuit board ALMEMO
(c) Measuring inputs M0 to M8 
      M0 to M8 for all ALMEMO sensors 
      M9 to M39 31 additional channels
(d)Code switches 
      G: Device address 0 to 99
(e)Key ON/OFF, START/STOP No function
(f) Output sockets A1, A2 
      A1 Interface / optic fiber (ZA1909-DK5/L) 
      RS 422 (ZA 5099-NVL/NVB) 
      Ethernet (ZA 1945-DK) 
      Bluetooth (ZA 1709-BTx)
A2 Network cable (ZA1999-NK5/NKL)
A1/A2 Trigger input (ZA 1000-ET/EK) 
      Relay outputs (ZA 1000-EGK) 
      Analog output 2 (ZA 1601-RK)
(g) Connection socket DC 12V 
      AC adapter (ZB 1212-NA3/USA, 12V, 2A) 
      Cable, electr. isol. (ZB 3090-UK2, 10-30V)
(h) Ground socket
(i) Check lamps 
      ON Device is on.
      START Measuring operation started
      REC Measuring with results saved
      COM Measuring with output
      ALARM Limit value exceeded

ORDERING INFORMATION

PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS
Model Description
Includes measuring instrument ALMEMO® 5690-2M, multimedia card and USB card reader, AC adapter ZB 1212-NA3/USA 12V, 2A, operating instructions, ALMEMO® Manual & CD with the AMR-Control software. 9 Inputs, max. 40 Channels, 2 Outputs, cascaded Interface, real time clock, 1 key
MA56902M09TG3 19“ desk-top housing 32 DU, 6 free slots
MA56902M09TG8 19“ desk-top housing 84 DU, 19 free slots
MA56902M09BT8 19“ desk-top housing 84 DU, 19 free slots
Options
ZB 1212-NA3/USA AC adapter, 12V, 2A
ZB3090UK2 C adapter cable, 10 to 30 VDC, 12 V / 1.25 mA, electr. isolated
OA5690S Options S: Internal EEPROM with capacity for 100,000 measures values
OA5690KL KL: Linearization, multi-point calibration, and calibration data management
SB0000R R: Temperature ranges for 8 coolants
ES5690AP Rechargeable battery pack (8 cells NiMH, 1600mAh), 1 free slot with intelligent high-speed charging (3.5h) Selector switch boards
ES5690UA10 U-A10 10 inputs for ALMEMO® flat connectors
ES5690UMU U-MU 10 inputs with 10x MU connector
ZA5690MU ALMEMO® 10-fold MU connector for 10 sensors, 10-40 channels
ES5690UTH U-TH 10 inputs with miniature thermal connector
ES5690UKS U-KS 10 inputs with clamp connector
OA5690UKSU Option KSU for ES5690UKS: inputs with 100:1 divider for 10 V
OA5690UKSI Option KSI for ES5690UKS : inputs with shunt for 20 mA
Accessories
ZB1904MMC Multimedia card min. 32 MB
ZA1909DK5/L ALMEMO® RS232data cable, optic fiber, max. 115.2kbaud, 1.5m
ZA1945DK Ethernet data cable, electr. isol., max. 115.2 KB, 1.5m
ZA1999NK5 Network cable, electr. isol., max. 57.6 KB, 1.5m
ZA5099-NVL RS422 network distributor, ALMEMO®, device connection via optic fiber, 1.5m
ZA5099-NTL RS232/RS422 network driver, ALMEMO®, computer connection via optic fiber, 1.5m
ZA1000-ET Trigger input cable for external triggering with key, trigger input 2-30 V optocoupler, 3mA, 1.5m
ZA1000EGK Trigger and alarm cable (2 relays, 1 ohm, 300 mA, 50 V)
ZA1601-RK Analog output cable, -1.25 to 2.000V, current cons. approx. 3 mA, >100 Ohm load, ±0.1%±6 digit accuracy

For assistance call 978-568-3400

pumps | valves | instrumentation | new products | contact clark | site/product directory | enewsletter | home