Model MA5690-2 data acquisition system with graphics display accepts 9 to 99 measuring inputs depending on the configuration ordered. Operation is easy by means of 4 soft keys and cursor block, comprehensive menu system with wizards and help windows.
- Ideal display with large, brightly illuminated
graphics
- Easy and convenient operation by means of
4 soft-keys and cursor block,
comprehensive menu system with wizards and help windows
- Choice of languages : German, English, French (others also available)
- 9 measuring menus (3 can be freely configured by user from a range
of 50 functions)
- Measured values can be displayed graphically in line chart or bar chart form
and from 1 to 20 measured values can be displayed numerically in various
sizes.
- 8 programming menus for easy-to-understand parametrization of cycles,
times, memory, device locking, output modules, and power supply, Sensor
programming with range, units, comments, scaling, error correction, etc.
- 9 LEDs for indicating various operating states
- Data logger with multimedia card, provided as standard, recording in standard
FAT16 file format, transmission to PC via USB card reader
- Option EEPROM with capacity for 100,000 measures values, internally
configurable as linear or ring memory
- Sleep mode for long-term recording
- Measuring functions Measured value, zero-setting, setpoint adjustment,
maximum and minimum values stored with date and time-of-day, smoothing,
average values over time or over measuring points, limit value monitoring,
cold junction compensation, temperature compensation, and atmospheric
pressure compensation
- Sensor programming on the ALMEMO® connector itself
- 9 ALMEMO® input sockets, electrically isolated, for 9 ALMEMO®
sensors, 31additional channels (maximum), (of which 4 internal
function channels)
- Extendible up to 99 inputs with various selector switch boards and maximum
100 meas. channels
- High-speed, high-resolution A/D converter, 24-bit, 2.5 to 100 measuring
operations per second
- Electr. isol. between measuring inputs and supply
- 2 ALMEMO® output sockets for digital interfaces, analog outputs, trigger input,
and alarm contacts
- 5 LEDs for indicating various operating states in the measuring circuit
- Cold junction measuring with 2 NTCs / insert and interpolation Ranges Pt100,
0.000 to 65.000 °C, timer 6500.0 seconds
Display: Graphics 128 x 128 dots, 16 rows, Lighting: 5 LEDs, 3 brightness levels Operation: 9 keys (4 soft-keys and cursor block), 9 status LEDs on the front panel Memory: Multimedia card and USB card reader Memory (option S): Internal 512-KB EEPROM (100,000 measured values) Date and time-of-day: Real-time clock, buffered with lithium battery
Master meas. board: 9 ALMEMO®input sockets, electr. isol., with semiconductor relays Channels: 9 primary channels, 27 additional channels for double sensors and function channels, 4 internal function channels A/D converter: delta-sigma, 24-bit, 2.5 / 10 / 50 / 100 measuring operations per second, electrically isolated Sensor power supply: AC adapter, 12 V, maximum 400 mA, Rechargeable battery, 9 to 11 V, maximum 200 mA
ES 5690-UA10: 10 electrically isolated inputs for ALMEMO® flat connectors, 10 to 40 channels, with sensor supply, 2 plug-in slots ES 5690-UMU: 10 electrically isolated inputs, sensor connection with 10x MU connector, 10 to 40 channels, without sensor supply, ranges only, for all thermocouples, Pt100, Ni100, NTC, ohms, 2.6 V, 260 mV, 55 mV, 26 mV, 1 plug-in slot ES 5690-UTH: 10 electr. isol. inputs for all thermocouples with miniature thermal connector, 10 to 40 channels,1+1 plug-in slots (dummy panel needed) ES 5690-UKS: 10 electr. isol. inputs, sensor connection with clamp connector, 10 to 40 channels, without sensor supply, ranges only, Pt100, Ni100, NTC, ohms, 2.6 V, 260 mV, 55 mV, 26 mV, 1 plug-in slot OA 5690-UKSU: Option for ES5690-UKS, inputs with 100:1 divider for 10 V OA 5690-UKSI: Option for ES5690-UKS, inputs with shunt for 20 mA ranges mA and (4-20 mA) Dimensions: TG 32 DU 179 x 158 x 232 mm (WxHxD) 6 slots TG 84 DU 444 x 158 x 232 mm (WxHxD) 19 slots BT 84 DU 483 x 132 x 273 mm (WxHxD) 19 slots Outputs: 2 ALMEMO®-sockets for all output modules (analog, data, trigger, relais cable etc.) internal alarm transmitter Power supply: AC adapter: ZB1212NA3/USA 90..250V, 12V DC, 2A; DC adapter cable, electr. isolated ZB3090-UK2, 10..30V,12V DC ,1.25A; Rechargeable battery pack 8 NiMH-Akkus 9..11V, 1600mAh
![]() |
(1) LCD
Status bar
Cont. Measuring point scan
Start / stop measuring
REC Record to memory
COM Measured value output
l>, >l Program start / end of measuring
R01 Status of alarm relays
* , * Lighting on, pause
Battery operation / charge status
13 lines for functions
Function of keys F1, F2, F3, F4
(2) Check lamps
START Measuring operation started
REC Measuring with results saved
COM Measuring with output
AVG Averaging
ALARM Limit value exceeded
Sensor breakage, LoBat
LOCKED Keys locked
SLEEP Flashes in sleep mode.
CHARGE Battery is being charged.
Goes out as soon as fully charged.
(3) Keypad
F1 to F4 Function keys (soft keys)
Cursor block
ON PROG Switch on Program, Switch OFF (press and hold down)
Arrow Keys: Function selection, input
(4) Slot, multimedia card: MMC slot for multimedia memory card |
![]() |
(5) Module AP rechargeable battery (option)
(a) Connection socket DC-A 12V
AC adapter (ZB 1212-NA3/USA, 12V, 2A)
(b) Check lamps
DC-A Mains supply present
CHARGE Battery is being charged.
Goes out as soon as fully charged.
(6) Module MM-A9 Measuring circuit board ALMEMO
(c) Measuring inputs M0 to M8
M0 to M8 for all ALMEMO sensors
M9 to M39 31 additional channels
(d)Code switches
G: Device address 0 to 99
(e)Key ON/OFF, START/STOP No function
(f) Output sockets A1, A2
A1 Interface / optic fiber (ZA1909-DK5/L)
RS 422 (ZA 5099-NVL/NVB)
Ethernet (ZA 1945-DK)
Bluetooth (ZA 1709-BTx)
A2 Network cable (ZA1999-NK5/NKL)
A1/A2 Trigger input (ZA 1000-ET/EK)
Relay outputs (ZA 1000-EGK)
Analog output 2 (ZA 1601-RK)
(g) Connection socket DC 12V
AC adapter (ZB 1212-NA3/USA, 12V, 2A)
Cable, electr. isol. (ZB 3090-UK2, 10-30V)
(h) Ground socket
(i) Check lamps
ON Device is on.
START Measuring operation started
REC Measuring with results saved
COM Measuring with output
ALARM Limit value exceeded |
| PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS | |
| Model | Description |
| Includes measuring instrument ALMEMO® 5690-2M, multimedia card and USB card reader, AC adapter ZB 1212-NA3/USA 12V, 2A, operating instructions, ALMEMO® Manual & CD with the AMR-Control software. 9 Inputs, max. 40 Channels, 2 Outputs, cascaded Interface, real time clock, 1 key | |
| MA56902M09TG3 | 19“ desk-top housing 32 DU, 6 free slots |
| MA56902M09TG8 | 19“ desk-top housing 84 DU, 19 free slots |
| MA56902M09BT8 | 19“ desk-top housing 84 DU, 19 free slots |
| Options | |
| ZB 1212-NA3/USA | AC adapter, 12V, 2A |
| ZB3090UK2 | C adapter cable, 10 to 30 VDC, 12 V / 1.25 mA, electr. isolated |
| OA5690S | Options S: Internal EEPROM with capacity for 100,000 measures values |
| OA5690KL | KL: Linearization, multi-point calibration, and calibration data management |
| SB0000R | R: Temperature ranges for 8 coolants |
| ES5690AP | Rechargeable battery pack (8 cells NiMH, 1600mAh), 1 free slot with intelligent high-speed charging (3.5h) Selector switch boards |
| ES5690UA10 | U-A10 10 inputs for ALMEMO® flat connectors |
| ES5690UMU | U-MU 10 inputs with 10x MU connector |
| ZA5690MU | ALMEMO® 10-fold MU connector for 10 sensors, 10-40 channels |
| ES5690UTH | U-TH 10 inputs with miniature thermal connector |
| ES5690UKS | U-KS 10 inputs with clamp connector |
| OA5690UKSU | Option KSU for ES5690UKS: inputs with 100:1 divider for 10 V |
| OA5690UKSI | Option KSI for ES5690UKS : inputs with shunt for 20 mA |
| Accessories | |
| ZB1904MMC | Multimedia card min. 32 MB |
| ZA1909DK5/L | ALMEMO® RS232data cable, optic fiber, max. 115.2kbaud, 1.5m |
| ZA1945DK | Ethernet data cable, electr. isol., max. 115.2 KB, 1.5m |
| ZA1999NK5 | Network cable, electr. isol., max. 57.6 KB, 1.5m |
| ZA5099-NVL | RS422 network distributor, ALMEMO®, device connection via optic fiber, 1.5m |
| ZA5099-NTL | RS232/RS422 network driver, ALMEMO®, computer connection via optic fiber, 1.5m |
| ZA1000-ET | Trigger input cable for external triggering with key, trigger input 2-30 V optocoupler, 3mA, 1.5m |
| ZA1000EGK | Trigger and alarm cable (2 relays, 1 ohm, 300 mA, 50 V) |
| ZA1601-RK | Analog output cable, -1.25 to 2.000V, current cons. approx. 3 mA, >100 Ohm load, ±0.1%±6 digit accuracy |