Get a printer friendly Acrobat™ file for each of the following:

AMR

MA8390-1 Universal Transmitter for ALMEMO®Sensors

ALMEMO Sensor Input, Ethernet, RS232 &
Analog Output

DESCRIPTION

Model MA8390 is a universal transmitter for all ALMEMO® sensors. It is especially easy and convenient to operate. The sensor programming is saved in the ALMEMO® connector. The transmitter automatically recognizes the sensor that is connected together with measuring range, units, comments, adjustment and correction values, and analog output scaling.

Transmitter and sensor can be changed quickly and easily. There is no need to alter the configuration. For customer-specific sensors, ready-fit ALMEMO® connectors are available.

Digital output is included as standard. The unit is ideally suitable for networked data acquisition and computer-supported on line operation. Sensors can be programmed and device parameters can be configured via interface commands. Packaging is a compact case with wall mounting or adapter for DIN rail mounting.

SPECIFICATIONS

Measuring Inputs: 1 ALMEMO® input socket for 1 ALMEMO® sensor
Channels: Per sensor max. four channels (sensor-type specific,
   measuring and function channels)
Input current: < 20 nA
A/D converter: Delta-sigma, 16-bit
Measuring ranges: as on Sensors for AMR Dataloggers bulletin
   but:
   Pt100 : with resolution 0.01K to +300°C
   Ohm : 0 to 500 ohms, resolution 0.1
   Volt DC : -2.0 to +2.6 V
   Milliampere DC : -26 to +26 mA
   Pt1000: not available
   Measuring current Pt100 : < 0.5 mA
System accuracy: ±0.03% of measured value ±3 digits
Sensor power supply Network adapter : approx. 12 V, maximum 70 mA
Outputs:1 ALMEMO® socket for analog output cable or for
   ALMEMO® data cable RS-232 / network cable
   Option E : Ethernet interface, electrically isolated,
   socket : RJ45 10/100 base-T, automatic switchover
   Option Rx :analog output R2 : 0-10 V; R3 : 4-20 mA electrically
   isolated, resolution : 15 bit
Standard equipment:
   Date and time-of-day: software clock (not buffered)
   Device address: 2 coding switches on the rear of the device
Power supply: 7 to 13 V DC, not electr. isolated
Option U : 10 to 30 V DC, electr. isolated
Current consumption: without input and output modules, without
   options approx. 5.5 mA
Housing: (HxWxD) 60 x 108 x 29 mm, ABS (acrylo-nitrile butadiene
   styrene) high impact resistance (max. 70°C)

FEATURES

ORDERING INFORMATION

FOR YOUR CONVENIENCE, THESE ITEMS ARE AVAILABLE ONLINE
Model Description
MA8390-1 Universal transmitter ALMEMO® 8390-1, Operating instructions,ALMEMO® Manual and AMR-Control software options
 
Options
OA8390-R3 Single analog output 4 to 20 mA, electr. isolated
 
Accessories
ZA1909-DK5 ALMEMO®RS232 data cable, electr. isolated, 9600 baud, 1 mA, 1.5m
ZB8390-H Mounting plate with 2 clips
ZB1012-NA2U/USA AC Power adapter 12 V / 200 mA
ZA1601-RK ALMEMO®analog output cable, -1.25 to 2.00 V, 0.1 mV / digit, 1.5m

Qty:

Show Discount Schedule

PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS
Model Description
MA8390-1 Universal transmitter ALMEMO® 8390-1, Operating instructions,ALMEMO® Manual and AMR-Control software options
 
Options
OA8390-R3 Single analog output 4 to 20 mA, electr. isolated
OA8390-U Power supply, electrically isolated, 10 to 30 V
OA8390-E DC Ethernet interface (OA8390U not possible)
OA8390-R2 Single analog output 0 to 10 V, electr. isolated
OA2290-HS Option HS : Fixture for DIN rail mounting
 
Accessories
ZA1909-DK5 ALMEMO®RS232 data cable, electr. isolated, 9600 baud, 1 mA, 1.5m
ZB8390-H Mounting plate with 2 clips
ZB1012-NA2U/USA AC Power adapter 12 V / 200 mA
ZA1601-RK ALMEMO®analog output cable, -1.25 to 2.00 V, 0.1 mV / digit, 1.5m
ZA5099NVL ALMEMO®network distributor RS-422 (see ALMEMO Network Connections bulletin)

 

   For assistance call 978-568-3400


pumps | valves | instrumentation | new products | contact clark | site/product directory | enewsletter | home