The model MA8590-9 is a 9-input data logger with internal memory capacity to store 100,000 measured values. Start and stop functions are accessed directly from the face of the unit. All other functions are accessed via PC connection. The unit may also be used to acquire data and deliver it directly to a PC. The model MA8590-9A is like the MA8990-8 but with rechargeable battery pack.
The MA8590-9 and -9A use the patented ALMEMO® connection system, whereby sensor-specific parameters are stored in non- volatile memory in the sensor's connection plug. This allows the user to make zero or span calibrations via digital interface through a PC, and the changes are stored automatically in the sensor's plug and remain with the sensor!
TECHNICAL FEATURES
- 9 ALMEMO®input sockets, electrically isolated, for 9 ALMEMO® sensors, 36 sensor channels, 4 additional internal function channels.
- 2 ALMEMO® output sockets for digital interfaces, analog output, trigger input, alarm contacts, memory card.
- High-speed, high-resolution A/D converter, 24-bit, 50 measuring operations per second. Electrically isolated between measuring inputs and supply.
- Programming of sensor parameters in the connector, can be modified via interface using the supplied AMR-Control software.
- Measuring functions: Measured value, zero-setting, sensor adjustment, maximum / minimum values stored with date, time-of-day, smoothing, average values over time or measuring points, limit value monitoring, temperature compensation, atmospheric pressure compensation.
- Cold junction measuring with 2 NTCs and interpolation. Ranges: Pt100 0.000 to 65.000°C, timer 6500.0 seconds.
- Option KL Multi-point calibration, calibration data management, user-defined linearization.
- 5 LEDs for indicating various operating states.
- EEPROM with capacity for 100,000 measured values, internally configurable as linear or ring memory.
- Sleep mode for long-term recording.
- Device programming using the supplied AMR-Control software, i.e. date and time-of-day, cycle, start and end of measuring, measuring rate, etc.
- Key for switching on, start / stop measuring.
- Variant with rechargeable battery system, 8 AA-type NiMH batteries, high-speed charging.
- Device software update via interface.
- Trigger variants, process control using command macros.
SPECIFICATIONS
Measuring inputs: 9 ALMEMO® input sockets, electr. isol., with semiconductor relays (50 V)
Channels: 9 primary channels, max. 32 additional channels for double sensors and function
channels (e.g. differential values)
A/D converter: Delta-sigma, 24-bit, 50 measuring operations per second, electr. isolated
Sensor power supply: AC adapter, 12 V, max. 150 mA Rechargeable bat., 9 to 12 V, max. 100 mA
Outputs: 2 ALMEMO® sockets for all output modules (analog, data, trigger, relay cables, memory etc.)
Standard equipment:
Operation: 1 key
Memory (Option S): Internal 512-KB EEPROM (100,000 meas. val.)
Date and time-of-day: Real-time clock, buffered with lith. battery
Power supply:
AC adapter: For model MA8590-9, 115 VAC to 12 VDC, electrically isolated 600 mA, P/N ZB1212NA4-USA;
For model MA 8590-9A,115 VAC to 12 VDC, electrically isolated 5A, ZB1212NA6-USA
Rechargeable battery: 8 NiMH rechargeable batteries, 1600mAh pack (8690-9A only) 2.5-hour
high-speed charging, internal
Current consumption without input and output modules:
Active mode approx. 25 mA
Sleep mode approx. 0.05 mA
Housing:
8590-9 : (LxWxH) 180 x 49 x 137 mm, polystyrene Weight 490 gr
8690-9A : (LxWxH) 218 x 77 x 145 mm, polystyrene
ORDERING INFORMATION
| FOR YOUR CONVENIENCE, THESE ITEMS ARE AVAILABLE ONLINE |
|
Model |
Description |
|
MA8590-9-OPTS |
9 inputs, 2 outputs, real-time
clock,AC adapter 600 mA, with Option S Internal Memory, 512-KB |
|
MA8590-9A-OPTS |
9 inputs, 2 outputs, real-time
clock, with rechargeable battery pack and AC adapter for charging, 5A and
with Option S Internal Memory, 512-KB |
|
ZA1601RK |
Analog output cable, -1.25 to 2.0 V, 0.1 mV / digit |
|
ZA1000EGK |
Trigger and alarm cable (2 relays, 1 Ohm, 300 mA, 50 V) |
|
ZA1945DK |
Ethernet data cable, electr. isol., maximum 115.2 KB |
|
ZA1999NK5 |
Network cable, electr. isol.,
maximum 115.2 KB |
| ZK1909DK5 |
V24 data cable, electr. isol., maximum 115.2 KB |
| OR PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS |
| Model |
Description |
| MA8590-9-OPTS |
9 inputs, 2 outputs, real-time clock, AC adapter 600 mA, with Option S Internal Memory, 512-KB |
| MA8590-9A-OPTS |
9 inputs, 2 outputs, real-time clock, with rechargeable battery pack and AC adapter for charging, 5A and with Option S Internal Memory, 512-KB |
| OA8590KL |
Multi-point calibration,linearization |
| OA2290HS |
Option top hat rail mounting |
| ZA1601RK |
Analog output cable, -1.25 to 2.0 V, 0.1 mV / digit |
| ZA1000EGK |
Trigger and alarm cable (2 relays, 1 Ohm, 300 mA, 50 V) |
| ZK1909DK5 |
V24 data cable, electr. isol., maximum 115.2 KB |
| ZA1945DK |
Ethernet data cable, electr. isol., maximum 115.2 KB |
| ZA1999NK5 |
Network cable, electr. isol., maximum 115.2 KB |