| Channel
switching between input sockets: |
4-contact with
photo-MOS relay |
| Potential
separation: |
Maximum 50 V
(for measuring modules with higher potential separation) |
| Offset
voltage: |
<5 μV |
| Measuring
rate: |
2.5 or 10
measuring operations per second |
| Cold junction
compensation: |
Effective in
range -30 to +100 °C Accuracy ±0.2 K ±0.01 K / °C |
| Nominal
temperature: |
22°C ±2 K |
| Sensor power
supply: |
7.2 to 12 V
unregulated, depending on power supply, maximum 70 mA/device |
|
Self-calibration: |
Automatic
zero-point correction, measuring current calibration |
| Check
functions: |
Automatic
sensor and sensor breakage detection |
|
Measuring rate: |
2.5 / 10 / 50 measuring operations per second |
|
Common-mode input range: |
-3 to +3 V in DC range (2.6 V) |
| -2.0 to +1.7 V in all other measuring ranges |
|
Overload maximum: |
±12 V |
|
Measuring current: |
Pt100 approx. 1 mA |
|
Pt1000 approx. 0.1 mA |
|
Input current: |
500 nA in DC range (2.6 V) |
|
500 pA in all other measuring ranges |
|
Functional restrictions at: |
analog output, connector scanning, sensor breakage detection |
| 50
meas. op. per second: |
alarm value printout, mains hum suppression |
|
System accuracy: |
0.02% ±1 digit at 2.5 und 10 measuring operations per second |
|
0.05% ±3 digit at 50 measuring operations per second |
|
Temperature drift: |
0.003 % / K |
|
ALMEMO® socket A1 |
Digital interface |
Baud rates:
150, 300, 600, 1200, 2400, 4800, 9600, 57.6K, 115.2K |
| Data: 8
bit serial, 1 start bit, 1 stop bit, no parity |
| RS232 with
data cable ZA1909DK5 |
| USB with
adapter cable ZB1909USB |
| Wireless with
Bluetooth modules ZA1709BT2DKU |
| RS422 with
branch box ZA5099NVL |
| Ethernet with
data cable ZA1945DK |
| Analog output
|
-1.25 to +2.0 V with recording cable ZA1601RK |
|
ALMEMO® socket A2 |
Networking |
Current loop
with network cable ZA1999NK5 |
| Wireless
with Bluetooth modules ZA1709BT2NK |
|
Data storage |
ALMEMO®
storage connector 128/256kB ZA 1904-SS |
| ALMEMO® memory
connector for multi-media card ZA1904MMC |
|
Analog output |
-1.25 to +2.0 V, not electr. isol., with recording cable ZA1601RK |
| -6 to +10 V, 0/4 to 20 mA, electr. isol., with adapter ZA8000RTA |
| Trigger input
|
With
trigger cable ZA1000-ET/EK/EAK, ZA8000RTA |
| Relay output |
With
relay cable ZA1000-EGK/EAK, ZA8000RTA |