INSTRUMENTATION > FLOW > DS07 Viscosity Compensated Variable Area Flowmeter and Switch, Mounting Independent
CLARK SOLUTIONS

DS07 Viscosity Compensated Variable Area Flowmeter and Switch, Mounting Independent


DESCRIPTION
The flow meter and switch model DS07 works according to a modified variable area principle. The float is guided in a cylindrical measuring glass by means of a spring. The flowing medium moves the float in the flow direction. The upper edge of the float shows the momentary flow via a burnt-in scale on the measuring glass.

A Reed contact is mounted outside the meter in a sealed housing. When the float reaches the position of the reed contact the switch will close. With higher flows the float moves further upward until it reaches a built-in float stop, still keeping the switch closed. This ensures a bistable switch function at any time. The Reed contact is adjustable over the full measuring range of the meter.

VISCOSITY COMPENSATION, MOUNTING POSITION AND RELIABILITY
The built-in spring and the magnetic float guarantee an absolute reliability of the meter. This spring, which pushes the float back towards its zero position against the flow makes it possible to use the meter in any mounting position. The spring is artificially aged, thus eliminating the need for recalibration to the different mounting positions. The strong spring and an orifice in the float work together to limit the effects of viscosity changes to an absolute minimum compared to regular variable area flowmeters.

TECHNICAL SPECIFICATIONS
Max Pressure:
  232 psi / 16 bar (DS07.M)
  145 psi / 10 bar (DS07.S)
Pressure Drop:
  0.29 - 2.9 psi / 0.02 - 0.2 bar (DS07.M)
  0.29 - 5.8 psi / 0.02 - 0.4 bar (DS07.S)
Max Temperature: 248°F/120°C
  (320°F/160°C on request)
Materials:
  Measuring glass: Duran 50
  Housing: anodized aluminium
  O-rings: Perbunan
               (optionally: Viton, EPDM)
Electrical Connections: plug acc. to DIN 43650
               (optionally: 1m cable connection) Accuracy: ±10% f.s.
Analog Output: see model DSxx-A in section
   "accessory"

DS07 Flow Meter/Switch
FEATURES

For Viscous Media up to 600 cSt


Mounts in any Position without Recalibration

Compact Design

Materials: Brass or Stainless Steel

High Switching Accuracy

Very Small Switch Hysteresis


Measuring Glass with Burnt-in Scale


APPLICATION
The variable area flow meter and switch model DS07 is used for measuring and monitoring the flow of viscous liquids, i.e., in central lubricating systems, any other lubricating circuitry, hydraulics, transformer oils, etc.

MEASURING RANGES
8-27 GPH … 8-24 GPM
0.2 - 0.8 l/min … 30 - 90 l/min
for viscosities up to max. 600 cSt

MATERIALS
brass or stainless steel
CONTACTS
**N/O: 250 V, 3 A, 100 VA
***SPDT: 250 V, 1.5 A, 50 VA
*Ex-N/O: 250 V, 2 A, 60 VA
*Ex-SPDT: 250 V, 1 A, 30 VA
 
* according to Atex 100a Ex II 2 G, EEx m II T6 and II 2D IP67 T80 °C
** for DS07.S.2/3/4...(230V, 1 A, 50 VA)
*** 250V, 1A, 50 VA (TYPE:2X)
DIMENSIONS:
Model Mounting dimensions in inch / mm Weight (lbs / g)
SW D B  NPT / G T L
DS07.S.1 1.61/41 1.97/50 3.03/77 1/4" 0.67/17 5.71/145 1.87/850
DS07.S.2 1.61/41 1.97/50 3.03/77 1/2" 0.67/17 5.71/145 1.87/850
DS07.M.1 1.06/27   2.07/53 1/2" 0.55/14 4.49/114 0.66/300
DS07.S.3 1.61/41 1.97/50 3.03/77 3/4" 0.67/17 5.47/139 1.87/850
DS07.S.4 1.61/41 1.97/50 3.03/77 1" 0.67/17 6.22/158 1.87/850
DS07 Diagram
Electrical Connection
ORDERING INFORMATION
Order Number Example: DS07. M. 2. 1. 1. 05. 1. 1. 0
DS07 Viscosity Compensated Variable Area Flowmeter and Switch, Mounting Independent
PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS
Model Size Connection Material Scale Measuring ranges No. of contacts Contact function
DS07.M 1/2" only: DS07.S 1/4" only DS07.S 1/2", 3/4", 1" DS07.S 3/4", 1"
DS07 M=miniature 1N=1/4" NPT female 1=brass, spring st. steel 1.4310 1=for viscous media 01U = - 06AU = - 06U = - 13U=160-475 GPH 0=without contact 0=without contact
S=standard 2N=1/2" female 02U = - 07AU=8-24 GPH 07U=8-24 GPH 14U=240-710 GPH 1=1 contact 1=N/O
  3N=3/4" female 2=all st. steel 1.4571 03U=8-27 GPH 08AU=16-63 GPH 08U=16-63 GPH 15U=320-950 GPH 2=2 contacts 2=SPDT
4N=1" female   04U=21-63 GPH 06 A=0.1-0.8 l/min 09U=32-127 GPH 16U=8-24 GPM   3S=Ex-N/O, not available for DS07.M (EEx m II T6)
1=G 1/4 female 05U=40-127 GPH 07 A=0.5-1.5 l/min 10U=48-159 GPH 13=10-30 l/min
2=G 1/2 female 01=0.2-0.8 l/min 08 A=1-4 l/min 11U=80-240 GPH 14=15-45 l/min
3=G 3/4 female 02=0.2-1 l/min   12U=125-380 GPH 15=20-60 l/min 3U=Ex-SPDT, not available for DS07.M (EEx m II T6)
4=G 1 female 03=0.5-1.7 l/min 06=0.1-0.8 l/min 16=30-90 l/min
  04=1.3-4 l/min 07=0.5-1.5 l/min  
05=2.5-8 l/min 08=1-4 l/min  
  09=2-8 l/min
10=3-10 l/min
11=5-15 l/min
12=8-24 l/min
 
Options
0 = without
1 = please indicate
This product sold by PKP Process Instruments, Inc., a joint venture of Clark Solutions
and PKP Prozessmesstechnik GmbH.