Clark Solutions FYA-Series Gas Concentration Sensors CO2, O2, CO, All ALMEMO Ready DESCRIPTION |
|||||||||||||||||||||||||||||||||
![]() |
||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||
![]() |
||||||||||||||||||||||||||||||||||
| SPECIFICATIONS CARBON DIOXIDE, FYA600-CO2 Measuring Principle: IR Optics Measuring Ranges: 0 0.5%, 0 2.5%, 0 10%, 0 25% Accuracy: +/-2%F.S. Linearity: +/-1%F.S. Current Consumption: nominal 50mA, maximum 70mA Settling Time (T90): <60s Operating Range: 5 40ºC, 0 95%RH, non-condensing Dimensions: W 96mm x H 36mm x D 64mm Weight: 241g Connecting Cable: 5 Feet OXYGEN, FYA600-O2 Measuring Principle: Electrochemical Cell Measuring Ranges: 1 100% Accuracy: +/-1% Resolution: 0.01% Response Time: <40s Signal Drift: <2% signal/month (typ. <5% over operating life) Operating Life: 2 years at 20.9% O2 Nominal Conditions: 20ºC, 50%RH, 1013mbar Operating Range: -20 50ºC, 0 99%RH, non-condensing Dimensions: H 43mm x Ø 29.3mm Weight: 241g Connecting Cable: 5 Feet CARBON MONOXIDE, FYA600-CO Measuring Principle: Electrochemical Reaction Measuring Range: 0 150ppm, 0 300ppm, 0 5% Zero Point Error: <10ppm CO Gauge Reading Balance: <3ppm CO Accuracy: +/-3%F.S. Zero Point Drift: <2% (1 year) Reproducibility: <2% (1 year) Linearity: <2%F.S. Settling Time (T90): <60s Operating Range: -10 +40ºC, 0 90%RH non-condensing Life Expectancy: approx. 2 years |
||||||||||||||||||||||||||||||||||