CLARK SOLUTIONS
Series 2000 Digital Process Indicator
Field Upgradeable Digital Process Indicators
Series 2000 Digital Process Indicator
DESCRIPTION
The Series 2000 Smart System Digital Indicator embodies many features and performance capabilities to suit a wide range of indication and control requirements. It can accept a variety of standard process signals and precisely scale them into any desired unit of measurement. The indicator employs advanced technology for stable, drift free readout, while incorporating features that provide flexibility now and in the future with plug-in option cards. The option cards afford the opportunity to easily configure the indicator for the needs of the present while providing an upward migration path as control and indication needs evolve. A full complement of options include relays, linearization, analog output and serial communication. All Series 2000 indicators are calibrated to your transducer requirements at no additional cost. Simply tell us how you want it set up, then plug it in and go; no muss, no fuss and no charge!
SPECIFICATIONS
Input Signals: Current, voltage or resistance
Process Display Keypad: 5 digit, 0.56 red LED, (-19999 to
       99999), 3 programmable function keys, 5 keys total
Power Requirement: 115/230 Vac or 11 Vdc to 36 Vdc  
Internal Power Supply: 24 Vdc
Electrical Connection: Terminal blocks in rear
Memory: Non-volatile Eprom, will hold set up data for 10
       years without power
Accuracy: ±0.03 % of reading +3 µA for 4 mA to 20 mA
       input; ±0.03 % of reading +3 mV for 0 Vdc to 5 Vdc
       and 0 Vdc to 10 Vdc inputs over the range of 18°C
       to 28°C
Update Rate: Up to 20 times per second, adjustable
Temperature Ranges: Operating: 32 °F to 122 °F
       (0°C to 50°C)
Storage: -40°F to 140°F (-40°C to 60°C)
Electromagnetic Rating: CE compliant to EMC norm EN
       61326:1997/A1:1998 RFI, EMI and ESD protection
Environmental Protection: Nema 4X/IP65 sealed bezel only
Linearization: 16 point scaling of non linear input
Dual Relay Option Card: Two FORM-C relays. Rating: One
       relay energized; 5 amps @120/240 Vac 115/230 Vac
       or 28 Vdc (resistive load), 1/8 HP @ 120 Vac, inductive
       load. Total current not to exceed 5 amps with both
       relays energized
Quad Relay Option Card: Four FORM-A relays. Rating: One
       relay energized; 3 amps @ 240 Vac or 30 Vdc (resistive
       load), 1/10 HP @ 120 Vac, inductive load. Total current
       not to exceed 4 amps with all four relays energized
Analog Output Option Card: 0 mA to 20 mA, 4 mA to 20 mA
       or 0 Vdc to10 Vdc retransmitted signal
Quad NPN-OC Option Card: 4 isolated open collector
       sinking transistors, 100 mA at maximum 30 Vdc
Quad PNP-OC Option Card: 4 isolated open collector
       sourcing transistors, 24 Vdc with 30 mA total maximum
Communication Option Cards: RS232C or RS485
Panel Cutout
ORDERING INFORMATION
ORDER NUMBER: ABCD
Example: 2000-231
PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS
A=Model B=Input Signal C=Power D=Option Cards
2000 1= 4-20 mA 1= 115/230 Vac 0= No Option Card
2= 0-5 Vdc 3= 11-36 Vdc 1= Dual Relay Card
4= 1-6 Vdc   2= Quad Relay Card
5= 0-10 Vdc 3= Quad NPN-OC Card
6= 1-11 Vdc 4= Quad PNP-OC Card
  5= Analog Output Card
6= RS 232-C Serial Communications Card
NEMA 4 Enclosure Option Card
For assistance call 978-568-3400