PRODUCT SPECIFICATIONS > INSTRUMENTATION
Series 110 Sanitary Pressure Transmitter Clark Solutions
Series 110 Sanitary Pressure Transmitter
Two & Three-wire, 4-20mA or Voltage output, Vacuum to 400 PSIG

DESCRIPTION
The Series 110 Sanitary Pressure Transmitter is designed for heavy duty sanitary applications where high accuracy and durability are required.  Using diffused semiconductor sensor technology, these transducers are stable, accurate, shock resistant and extremely durable.

The housing is constructed of 316SS and welded to the process connection for greater strength and integrity. The available 1-1/2 inch or 2 inch Tri-Clamp® connection, with its integral cooling extension, is 316L stainless steel and wetted parts are electro-polished to Ra25 micro-inch or better.

Series 110 Sanitary Transmitters meet 3A requirements for the food & beverage, dairy, pharmaceutical and biotechnology industries in addition to ASME BPE-2002 and are CE compliant.

A final electrical output and calibration inspection is performed on all transmitters prior to shipment to ensure 100% “out-of-the-box” reliability.

.

SPECIFICATIONS
Output Signals: 4-20 mA 2-wire; 0V-5 Vdc, 3-wire; 1-5 Vdc,
      3-wire; 1-6 Vdc, 3-wire; 0-10 Vdc, 3-wire; 1-11 Vdc, 3-wire
Pressure Ranges: Standard gauge ranges from vacuum to
      400 PSIG
Proof Pressure: 3 times Full Scale for ranges 0-2 PSIG through
      0-200 PSIG; 1.75 times
      Full Scale for ranges 0-300 PSIG through 0-400 PSIG
Burst Pressure: 3.8 times Full Scale for ranges 0-2 PSIG
      through 0-200 PSIG; 4 times
      Full Scale for ranges 0-300 PSIG through 0-400 PSIG
Accuracy: ±0.25% Full Scale (B.F.S.L), ±0.125%
      Full Scale (optional)
Repeatability: ±0.05% Full Scale
Hysteresis: ±0.1% Full Scale
Stability: ±0.2% Full Scale for 1 year, non-accumulating
Power Supply: 10-30Vdc for current output 14-30Vdc for
      voltage output
Case Materials: 316 stainless steel
Temperature Ranges: Compensated 32ºF to 175ºF
      (0ºC to 80ºC)
Effect: ±0.01%/ºF for zero and span
Ambient: -40ºF to 176ºF (-40ºC to 80ºC)
Adjustment: ±10% Full Scale for zero and span
Environment Protection: NEMA 4X, IP65 (IEC 529)
Electromagnetic Rating: CE compliant to EMC norm
EN61326: 1997/A1: 1998 RFI, EMI, ESD protection
Electrical Protection: Reverse polarity, overvoltage and short
      circuit protection
Process Connection: 1 1/2 inch or 2 inch Tri-Clamp®
Seal Housing Material: 316L stainless steel
Diaphragm Material: 316L stainless steel electropolished to
      Ra25 or better
Fill Fluid: White Oil (FFL 77), USP grade
Media Temperature: -40ºF to 300ºF (-40ºC to 150ºC)

Series 110 diagram
ORDERING INFORMATION
ORDER NUMBER ABCDEFGHI
EXAMPLE: 110-12461513411
PLEASE CALL 1-800-253-2497 TO ORDER THESE ITEMS
A=Model B=Clamp Size C=Seal Fill Fluid D=Inventory ID F=Accuracy G=Pressure Range H=Output Signal I=Electrical Connection
110 12=1-1/2" 4=FFL77 White Oil 615 1=±0.25% F.S. 40=0-15PSIG 1= 4-20mA 1=36" cable attached to Hirschmann
16=2" 2=±0.125% F.S. 43=0-30PSIG 8=Hirschmann (DIN 43650A)

 

 

 

 

 

 

46=0-60PSIG 14=1/2" ISO 4400 conduit
49=0-100PSIG 23=cable gland with internal junction box
52=0-150PSIG 29=1/2" NPT female conduit w/internal junction box
58=0-200PSIG 36=integral 36" cable