TECHNICAL BULLETIN

TAKASAGO ZERO INTERNAL VOLUME OPTION
Prevents Cross-Contamination, Reduces Sample And Reagent Volumes

TYPICAL APPLICATIONS
The internal volume is the space between the isolation diaphragm and the valve body. When the valve closes, this volume is trapped, often reducing the sensitivity of the instrument. In a conventional valve seat design, the protruding valve seat is hard and prevents the Teflon diaphragm from flattening against the valve body.

In the zero internal volume system, a special soft inert elastomer (perfluoroelastomer) is used for the valve seat. When the valve closes the soft protrusion presses flat, eliminating the internal volume.

FEATURES

CROSS - CONTAMINATION PREVENTION
FASTER FLUSHING
PERFECT EJECTION OF AIR BUBBLES
REDUCE USE OF SAMPLE AND REAGENTS

The zero internal volume option is available for 2-way valve model types STV and MTV.

Model Type Internal Volume
MTV Conventional 87mul
MTV Zero Internal Vol. *34mul
STV Conventional 87mul
STV Zero Internal Vol. *16mul
*Passage Volumes Only
Takasago Diaphragm Pumping Volume

Typical Pumping Volume mul
Model Port On Off
MTV-3 Way Comm. -0.64 -0.34
N.C. -0.64 +2.01
N.O. 0.00 -.034
ExakN-3 Way Comm. +0.015 -0.015
N.C. +0.007 0.000
N.O. 0.000 0.000