| PROBLEM |
POSSIBLE CAUSES |
SOLUTIONS |
| 1. Valves do not open when energized (NC) or when de-energized (NO). |
For direct acting valves |
|
1. Voltage less than 15% of the nominal voltage.
|
1. Check the coil voltage, which must not be less than 85% of the indicated nominal voltage. If this is the case, adjust the source to the adequate value. |
|
2. Too high a differential pressure for that model.
|
2. Reduce pressure to the maximum shown on the valve nameplate or change it for a more adequate one.
|
|
3. Burnt coil (with the circuit open).
|
3. See Burnt Coils (Problem 3).
|
|
4. Plunger jammed with solids.
|
4. Clean the plungers core tube and the valve. If the system lacks an adequate strainer before the valve, the problem will persist.
|
|
5. Damaged plunger.
|
5. Replace the damaged part. Damage may be caused by fluid abrasive elements or high operation frequency over a long period of time and exceeding the parts life.
|
|
For pilot operated valves
The same as above plus:
|
|
6. Differential pressure too low.
|
6. This factor should be considered when choosing a valve. It may occur due to over-sizing or reduction of differential pressure. If differential pressure cannot be increased by increasing the flow, the valve must be changed for an adequate one.
|
|
7. Jammed pilot piston.
|
7. Check that solids have not affected the pistons movement. After cleaning, check that it is not damaged. A strainer must be placed upstream from the valve to eliminate the problem.
|
|
8. Damaged pilot piston, pilot piston rings or diaphragm.
|
8. Change damaged parts. Check that the cause is not dirt. Also see solution #1.
|
|
9. Pilot orifice blocked.
|
9. Clean the orifice, if the orifice is damaged consult Clark. Also see solution #1.
|
|
10. Pilot gasket damaged or misaligned.
|
10. This is caused by poor assembly. Change the damaged part and assemble the valve correctly. The O-ring must be correctly fitted.
|
| 11. Excessive viscosity. |
11. Fluids with vicosities exceeding 60 cSt cannot be used with pilot operated valves |
| 2. The valve remains open |
For direct acting valves |
| 1. The coil was not de-energized (NC valve) or energized (NO valve). |
1. Check the control circuits. |
|
2. Plunger jammed with solids.
|
2. Clean the plungers core tube and the valve. If the system lacks an adequate strainer before the valve, the problem will persist.
|
For pilot assisted valves
The same as above plus: |
|
3. The pilot orifice does not close.
|
3. Check that the plunger is not jammed or the seats damaged. In the first case, clean it, in the second case, change it. If the orifice seat is damaged, consult Clark.
|
|
4. Compensation orifice blocked.
|
4. Clean the orifice, if the orifice is damaged consult Clark.
|
|
5. Jammed pilot piston.
|
5. Check that solids have not affected the pistons movement. After cleaning, check that it is not damaged. A strainer must be placed upstream from the valve to eliminate the problem.
|
| 6. Pilot piston, Pilot piston rings or diaphragm damaged. |
6. Change the damaged parts. Check that the cause is not dirt. |
| 7. Excessive viscosity. |
7. Fluids with viscosities exceeding 60 cSt cannot be used |
| 3. The coil gives off a burning smell after working for a short period or it burns up frequently. |
1. Excessive voltage.
|
1. The voltage must not exceed 10% of the nominal voltage, and only for brief periods. Correct the voltage. |
|
2. Only for AC: Too high a pressure that does not allow the pilot to open, therefore, only inrush current is present, which doubles the holding current.
|
2. Adjust the maximum working pressure to the maximum shown on the nameplate. If pressure is within the parameters, check that voltage is not less than 85% of the nominal voltage.
|
|
3. The coils nominal voltage is less than the sources or does not correspond to its cycling.
|
3. Check that the voltage and current type is as indicated on the coil.
|
| 4. Excessive fluid or ambient temperature. |
4. The fluid, atmosphere and power of the coil determine the internal temperature. As a general rule, the fluid temperature + ambient temperature must not exceed 210ºC. The fluid temperature cannot be above 180ºC. When handling hot fluids and the ambient exceeds 30ºC, it is advisable to fit the valve in the most ventilated area.
|
| 5. Moisture entering the interior of the coil |
5. Check that DIN coils strain relief is tight and the armored cable corresponds to the connector PG size. For series 1314, check that the housing and connection are closed. See mounting recommendations.
|
| 6. Lack of part of the electromagnetic package when it is not integrated to the coil. |
6. Replace the missing parts because they are part of the magnetic circuits and their absence results in an increase of the intensity which reduces the force of the magnetic attraction.
|
| 7. It is energized outside the valve (AC only). |
7. Do not energize the coil if it is not fitted to the valve. |
| 4. The coil vibrates when energized. |
1. Insufficient voltage. |
1. Adjust the voltage within the permitted parameters. |
| 2. Dirty fixed core and plunger surfaces, they have scales. |
2. Clean the surfaces. If scales remain there, change the components. |
| 5. Fluid leakage when closed. |
1. Main or pilot seat damaged or dirty. |
1. Clean or change seats. If the orifice seats are damaged, consult Clark. |
| 6. It operates slowly or fails. |
1. Compensation or pilot orifice partially blocked. |
1. In case of dirt, clean the orifices. In case of damage, consult Clark. |
| 2. Excessive fluid viscosity. |
2. The fluids viscosity must not exceed 60 cSt. |
| 3. Temporary excess or lack of differential pressure. |
3. Check that both differential and opening pressure differential are within the limits indicated in the valve nameplate. |